Test 2-8: Duty Cycle (cont’d)

Test Procedure
(cont’d)

4.

Set the AFG frequency range as specified in Table 2-8:

FREQ:RANG MAX

Enable doubling

or

FREQ:RANG MIN

Disable doubling

5.

Set AFG output frequency:

FREQ <frequency>

Set frequency

where <frequency> is the value specified in Table 2-8.

6.

Initiate the waveform:

INIT:IMM

7.

Measure positive pulse width (average at least 10 periods) with the
Counter and record the reading in Table 2-11.

NOTE

If a percentage result is desired, measure the period (average at least 10
periods. Duty Cycle (%) = 100 x (Positive Pulse Width/Period)

Table 2-8. Duty Cycle Test Points

Frequency

(Hz)

Frequency

Range

Test Limits

(sec)

1.0 E3
2.0 E3
2.5 E5
5.0 E5

MIN
MAX
MIN
MAX

5.0E-3

±

1.0E-6

2.5E-4

±

3.0E-5

2.0E-7

±

3.4E-9

1.0E-7

±

1.5E-8

Agilent E1445A Service Manual

Verification Tests 61