Test 2-9: Total Harmonic Distortion (cont’d)

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Set the AFG to output a sinewave with the 10 MHz filter enabled:

VOLT 24DBM

Set AFG amplitude

OUTP:FILT:FREQ 10 MHZ

Set filter to 10 MHz

OUTP:FILT ON

Enable filter

INIT:IMM

Initiate waveform

Perform steps 3 - 6 for each frequency listed in Table 2-9:

3.

Set AFG output frequency:

FREQ <frequency>

Set frequency

where <frequency> is the value specified in Table 2-9.

4.

Set the Spectrum Analyzer center frequency to the output frequency
of the AFG. Measure the peak amplitude of the fundamental in dBm.
Note the result for use in step 6.

5.

Set the Spectrum Analyzer center frequency to the second harmonic (
2

× fundamental frequency ). Measure the peak amplitude of the

second harmonic in dBm.

6.

Repeat step 5 for third through ninth harmonics. Calculate total
harmonic distortion as shown below:

thd

(

dBc

)

=

20

log

ж

и

√

result

2

2

+

result

3

2

+

+

result

9

2

ц

ш

where

result

n

(

volts

)

=

10

[

(

n

th

Harmonic

(

dBm

)

Fundamental

(

dBm

))

/20 ]

Agilent E1445A Service Manual

Verification Tests 65