Functional Verification: Gate In Test (cont’d)

Test Procedure
(cont’d)

4.

Set up the AFG to output a 1 MHz sinewave with an external gate
source:

TRIG:GATE:SOUR EXT;

External gate source

:TRIG:GATE:STAT ON;

Enable gate

:FREQ 1E6;

Set freq to 1 MHz

:VOLT 4VPP

Set AFG amplitude

INIT:IMM

Initiate

5.

Send the following command to the Command Module to set the
level at the "Trig Out" port to 5 V. Verify that the scope shows a 1
MHz sinewave.

OUTP:EXT:LEV 0

6.

Send the following command to the Command Module to set the
level at the "Trig Out" port to 0 V. Verify that the scope shows a DC
signal.

OUTP:EXT:LEV 1