Test 2-6: AC Flatness - 10 MHz Filter (cont’d)

Test Procedure
(cont’d)

7.

Set up the Power Meter:

Units - Watts
Power Range - auto
Reference Oscillator - ON

NOTE

Follow the Power Meter manufacturer’s instructions for performing an
autocalibration and correcting for the power sensor.

8.

Connect the equipment as shown in Figure 2-7:

9.

Set the Power Meter expected frequency to the crossover frequency
(100 kHz). Measure the AFG output power and convert the reading
to volts:

Reading

(

volts

)

=(

√

ï

Reading

(

watts

)

п

Ч

50

)

Figure 2-7. Equipment Setup for Test 2-6

Agilent E1445A Service Manual

Verification Tests 53