Agilent Technologies 75000 Series C User Manual

Page 82

Advertising
background image

Table 2-11. Performance Test Record for the Agilent E1445A (Page 6 of 7)

Model _____________________________ Report No. ____________________________ Date _______________

Test

Description

Minimum

Measured

Reading

Maximum

Meas

Uncert

TAR

Test 2-6. AC FlatnessTest - 10 MHz Filter (Values in dB error, relative to 1 kHz)

Amplitude at 24 dBm:

400 kHz
800 kHz
1.2 MHz
1.6 MHz
2.0 MHz
2.4 MHz
2.8 MHz
3.2 MHz
3.6 MHz
4.0 MHz
4.4 MHz
4.8 MHz
5.2 MHz
5.6 MHz
6.0 MHz
6.4 MHz
6.8 MHz
7.2 MHz
7.6 MHz
8.0 MHz
8.4 MHz
8.8 MHz
9.2 MHz
9.6 MHz
10.0 MHz
10.4 MHz
10.8 MHz

-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2
-0.2

____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________

0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2

0.0478 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0506 dB
0.0536 dB
0.0536 dB
0.0536 dB
0.0536 dB
0.0536 dB
0.0536 dB
0.0536 dB
0.0536 dB
0.0536 dB
0.0536 dB
0.0536 dB

4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1
4:1