AC Flatness Adjustment Procedure (cont’d)

Example Program (cont’d)

420 Flatness:SUB Flatness(Filter$,Mode$)
430 COM @Afg,@Dmm,@Pwr_mtr,@Analyzer,Secure_code$
440 COM /Flat/ INTEGER Num_points,Max_con
450 INTEGER Filter,Ac_cal_int(1:2)
460 CLEAR SCREEN
470 !
480 !---------- Initialize variables ----------
490 Ampl_dbm=24

!AFG max amplitude

500 !
510 !---------- Main Program ----------
520 IF Filter$="250KHZ" THEN
530 Num_points=25

!Number of test points

540 Max_freq=2.50E+5
550 ELSE

!Else, 10M filter will be used

560 Num_points=27

!Number of test points

570 Max_freq=1.08E+7
580 END IF
590 PRINT "FILTER = "&Filter$
600 !
610 ALLOCATE Test_freq(1:Num_points),Results(1:Num_points)
620 !
630 !Determine test frequencies
640 Step_size=Max_freq/Num_points
650 FOR I=1 TO Num_points
660 Test_freq(I)=Step_size*I
670 NEXT I
680 !
690 GOSUB Setup_afg
700 CALL Meas_flat(Test_freq(*),Results(*),Filter$)
710 !
720 IF Mode$="A" THEN
730 CALL Adj_flat(Results(*),Filter$,Test_freq(Num_points))
740 END IF
750 !
760 DEALLOCATE Test_freq(*),Results(*)
770 SUBEXIT
780 !
790 Setup_afg: !
800 OUTPUT @Afg;"*RST;*CLS"
810 WAIT .5
820 !

(Continued on next page)

Agilent E1445A Service Manual

Adjustments 97