Defeating calibration security – Agilent Technologies 75000 Series C User Manual

Page 88

Advertising
background image

Defeating Calibration Security

If the calibration security code is unknown, the security feature can be
defeated by disassembling the AFG and moving the jumper on connector
J104 (see Figure 3-1) to the unsecured position (left-most pins). To prevent
accidental or unauthorized calibration, move the jumper back to the secured
position (right-most pins) as soon as the security code has been set to the
desired value (use the CALibration:SECure:CODE <code> command).
Disassembly instructions can be found in Chapter 5.

Figure 3-1. Disabling Calibration Security

(shown in secured position)

86 Adjustments

Agilent E1445A Service Manual

Advertising
This manual is related to the following products: