DC Adjustment Procedure (cont’d)

Example Program (cont’d)

450 IF Cal_point=30 THEN

!Special case -- set range now

460 OUTPUT @Dmm;"RANGE 10"
470 END IF
480 !
490 OUTPUT @Afg;"CAL:DC:POINT? ";Reading

!Send cal value to AFG

500 ENTER @Afg;This_point,Err_num

!Returns current point,err code

510 WAIT .5
520 !
530 Results(Cal_point)=PROUND(Reading,-5)
540 Cal_point=Cal_point+1

!Increment Cal_point

550 UNTIL (Err_num<>0) OR (Cal_point>Max_cal_point)
560 !
570 !---------- End of loop ----------
580 !If error, print error number & cal point, else send PUD string
590 IF Err_num=0 THEN
600 PRINT "Calibration Successful"
610 !
620 !Store cal information if desired - place desired data inside quotes
630 !in following line and remove !’s.
640 ! Pud$="63 CHARACTERS MAX"

!Change Pud$ as desired

650 ! OUTPUT @Afg;"*PUD #0"&Pud$;CHR$(10);END
660 ELSE
670 PRINT "Calibration Error Number "&VAL$(Err_num)&" at Cal Point "&VAL$(Cal_point-1)
680 END IF
690 DISP
700 !
710 !---------- Quit ----------
720 PAUSE
730 OUTPUT @Afg;"CAL:SEC:STATE ON"

!Enable cal security

740 OUTPUT @Afg;"*RST"

!Reset AFG

750 OUTPUT @Dmm;"RESET"
760 LOCAL @Dmm

!Return DMM to local control

770 !
780 ASSIGN @Afg TO *
790 ASSIGN @Dmm TO *
800 STOP

!End of main program

810 !
820 !---------- Subroutines ----------
830 !
840 Setup_dmm: !
850 SELECT Cal_point
860 CASE =1

!Cal point 1

870 OUTPUT @Dmm;"FUNC DCV;RANGE 10;NPLC 10;OCOMP OFF"
880 CASE =29

!Cal point 29

890 OUTPUT @Dmm;"RANGE 1;NPLC 1"

(Continued on next page)

Agilent E1445A Service Manual

Adjustments 91