Table 2-11. Performance Test Record for the Agilent E1445A (Page 3 of 7)

Model _____________________________ Report No. ____________________________ Date _______________

Test

Description*

Minimum

Measured

Reading

Maximum

Meas

Uncert

TAR

Test 2-1. DC Zeros Test (Values in Vdc)

No Filter:

10.2375V (0 dB atten)
9.1347V (.99 dB atten)
9.1241V (1 dB atten)
8.1319V (2 dB atten)
6.4594V (4 dB atten)
4.0756V (8 dB atten)
2.2918V (13 dB atten)
2.0426V (14 dB atten)
0.3238V (30 dB atten)

-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.0044
-0.0044

____________
____________
____________
____________
____________
____________
____________
____________
____________

0.022
0.022
0.022
0.022
0.022
0.022
0.022
0.0044
0.0044

1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6

>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1

250 kHz Filter:

10.2375V (0 dB atten)
9.1347V (.99 dB atten)
9.1241V (1 dB atten)
8.1319V (2 dB atten)
6.4594V (4 dB atten)
4.0756V (8 dB atten)
2.2918V (13 dB atten)
2.0426V (14 dB atten)
0.3238V (30 dB atten)

-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.0044
-0.0044

____________
____________
____________
____________
____________
____________
____________
____________
____________

0.022
0.022
0.022
0.022
0.022
0.022
0.022
0.0044
0.0044

1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6

>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1

10 MHz Filter:

10.2375V (0 dB atten)
9.1347V (.99 dB atten)
9.1241V (1 dB atten)
8.1319V (2 dB atten)
6.4594V (4 dB atten)
4.0756V (8 dB atten)
2.2918V (13 dB atten)
2.0426V (14 dB atten)
0.3238V (30 dB atten)

-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.0044
-0.0044

____________
____________
____________
____________
____________
____________
____________
____________
____________

0.022
0.022
0.022
0.022
0.022
0.022
0.022
0.0044
0.0044

1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6

>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1

* Since the arbitrary waveform consists of zeros, the expected reading is 0 V, regardless of the amplitude
setting. The amplitude is changed in order to turn the various attenuators on and off.