Test 2-9: Total Harmonic Distortion (cont’d)

Example Program (cont’d)

230 PRINT "Connect Spectrum Analyzer to AFG Output."
240 DISP "Press ’Continue’ when ready"
250 PAUSE
260 !
270 !---------- Set up AFG ----------
280 OUTPUT @Afg;"*RST"

!Reset AFG

290 WAIT 1
300 OUTPUT @Afg;"VOLT "&Dbm_out$

!Set AFG output

310 OUTPUT @Afg;"OUTP:FILT:FREQ 10MHZ"

!Enable 10MHz filter

320 OUTPUT @Afg;"OUTP:FILT ON"
330 OUTPUT @Afg;"INIT:IMM"

!Initiate

340 !
350 !---------- Perform test ----------
360 FOR I=1 TO 5
370 OUTPUT @Afg;"FREQ "&VAL$(Freq(I))

!Set frequency

380 !
390 CALL Meas_thd(Freq(I),Thd)

!Measure THD

400 PRINT "Fundamental Frequency =";Freq(I)
410 PRINT "THD =";DROUND(Thd,4);"dBc"
420 PRINT
430 DISP "Press ’Continue’ when ready"
440 PAUSE
450 DISP
460 NEXT I
470 !
480 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

490 END
500 !
510 !
520 !---------- Measurement subprogram ----------
530 SUB Meas_thd(Frequency,Thd)
540 INTEGER Harmonic
550 CLEAR SCREEN
560 Harmonic=1
570 !
580 GOSUB Meas_fund

!Get fundamental amplitude

590 !

(Continued on next page)

Agilent E1445A Service Manual

Verification Tests 67