Test 2-10: Spurious/Non-Harmonic Distortion

Description

The purpose of this test is to verify that the AFG meets its specifications for
non-harmonic and spurious distortion.

Equipment Setup

Connect equipment as shown in Figure 2-9

Set Spectrum Analyzer to :

Ref Level = -5 dBm
Resolution BW = 3 kHz
Video BW = 3 kHz

NOTE

These are recommended settings only. Adjust your Spectrum Analyzer as
necessary.

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Set the AFG to output a -5 dBm, 10 MHz sinewave with the 10 MHz
filter enabled:

FREQ 1.0E7;

Set AFG frequency

:VOLT -5DBM

Set AFG amplitude

OUTP:FILT:FREQ 10 MHZ

Set filter to 10 MHz

OUTP:FILT ON

Enable filter

INIT:IMM

Initiate waveform

Agilent E1445A Service Manual

Verification Tests 69