Dc adjustment procedure – Agilent Technologies 75000 Series C User Manual

Page 89

Advertising
background image

DC Adjustment Procedure

Description

A DC adjustment is performed on the AFG by reading a series of voltages and
resistances output by the AFG, then entering those values back into the AFG.
After all measurements have been completed, new calibration constants are
calculated and stored in non-volatile memory. To ensure accuracy, perform the
DC calibration procedure at one year intervals.

This procedure uses a firmware routine to adjust the AFG’s DC calibration
constants. The CALibration[:DC]:BEGin command starts the DC calibration
sequence and the CALibration[:DC]:POINt? command steps the AFG to the
next calibration point. All AFG settings are performed by the firmware
routine.

NOTE

If an error occurs at any time during the procedure, abort (*RST) and start over.

Equipment Setup

Perform a complete autocalibration on the DMM (unless an
autocal has been performed within the last 24 hours)

Connect the equipment as shown in Figure 3-2

Figure 3-2. DC Adjustment Setup

Agilent E1445A Service Manual

Adjustments 87