Test 2-10: Spurious/Non-Harmonic Distortion (cont’d)

Test Procedure
(cont’d)

Perform steps 3 and 4 for each frequency range listed in Table 2-10:

3.

Set the Spectrum Analyzer start frequency and stop frequency to the
values listed in Table 2-10.

4.

Measure the amplitude (in dBm) of the highest peak. Subtract the
amplitude of the fundamental (-5dBm) from the reading and record
the result in Table 2-11:

result

(

dBc

)

=

reading

(

dBm

)

(−

5 dBm

)

Table 2-10. Spurious/Non-Harmonic Test Points

Start

Frequency

(Hz)

Stop

Frequency

(Hz)

Test Limits

(dBc)

100 E3
10.5 E6
21 E6
31 E6
41 E6
51 E6
75 E6
100 E6
125 E6

9.5 E6
19 E6
29 E6
39 E6
49 E6
75 E6
100 E6
125 E6
150 E6

-45
-45
-45
-45
-45
-45
-45
-45
-45

70 Verification Tests

Agilent E1445A Service Manual