Functional Verification: Output Relay Test

Description

The purpose of this test is to check the output relay.

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Set up equipment as shown in Figure 2-4:

3.

Set up the AFG to output a 1 MHz sinewave:

FREQ 1E6;

Set freq to 1 MHz

:VOLT 4VPP

Set AFG amplitude

INIT:IMM

Initiate

4.

Verify that a 1 MHz sinewave appears on the scope.

5.

Disable the Output relay:

OUTP OFF

6.

Verify that no signal appears on the scope.

Figure 2-4. Output Relay Test Setup

Agilent E1445A Service Manual

Verification Tests 25