Agilent Technologies 75000 Series C User Manual

Page 52

Advertising
background image

Test 2-5: AC Flatness - 250 kHz Filter (cont’d)

Test Procedure
(cont’d)

3.

Set the AFG output to the reference frequency (1 kHz):

FREQ 1000

Set frequency

4.

Measure the amplitude with the DMM and convert the reading to
dBm. Note the result for use in step 6:

Reference Level

(

dBm

)

=

20

×

log

ï

Reading

(

volts

)ï+

13.0103

Perform steps 5 - 6 for each frequency listed in Table 2-5:

5.

Set the AFG output:

FREQ <frequency>

Set frequency

where <frequency> is the value specified in Table 2-5.

6.

Measure the amplitude with the DMM and convert the reading to
dBm. Calculate and record the error relative to the reference level
calculated in step 4:

Reading

(

dBm

)

=

20

×

log

ï

Reading

(

volts

)ï+

13.0103

Error

(

dB

)

=

Reading

(

dBm

)

Reference Level

(

dBm

)

50 Verification Tests

Agilent E1445A Service Manual