Test 2-9: total harmonic distortion (cont’d) – Agilent Technologies 75000 Series C User Manual

Page 68

Advertising
background image

Test 2-9: Total Harmonic Distortion (cont’d)

Test Procedure
(cont’d)

Table 2-9. THD Test Points

Frequency

(Hz)

Test Limits*

(dBc)

100 E3
250 E3
1 E6
4 E6
10 E6

-60
-60
-48
-36
-36

* Through 9th harmonic

Example Program

This program performs the Total Harmonic Distortion Test.

10 ! RE-STORE "SINE_THD"
20 DIM Freq(1:5)
30 !
40 !---------- Set up I/O path and reset AFG ----------
50 ASSIGN @Afg TO 70910

!AFG I/O path

60 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

70 !
80 !---------- Initialize variables ----------
90 Dbm_out$="24DBM"

!AFG output

100 !
110 DATA 100E3,250E3,1E6,4E6,10E6

!Read in freqs

120 READ Freq(*)
130 !
140 !---------- Set up Spec Analyzer ----------
150 CLEAR SCREEN
160 PRINT "Set up Spectrum Analyzer:"
170 PRINT
180 PRINT " Ref Level = 25 dBm"
190 PRINT " Span = 1 kHz"
200 PRINT " Resolution BW = 30 Hz"
210 PRINT " Video BW = 30 Hz"
220 PRINT

(Continued on next page)

66 Verification Tests

Agilent E1445A Service Manual

Advertising
This manual is related to the following products: