Agilent Technologies 75000 Series C User Manual

Page 80

Advertising
background image

Table 2-11. Performance Test Record for the Agilent E1445A (Page 4 of 7)

Model _____________________________ Report No. ____________________________ Date _______________

Test

Description

Minimum

Measured

Reading

Maximum

Meas

Uncert

TAR

Test 2-2. DC Accuracy Test (Values in Vdc)

No Filter:

10.2375V
5.0V
0.0V
-5.0V
-10.24V

10.1863

4.9645

-0.0205
-5.0355

-10.2912

____________
____________
____________
____________
____________

10.2887

5.0355
0.0205

-4.9645

-10.1888

4.9E-5
2.5E-5
1.0E-6
2.5E-5
4.9E-5

>10:1
>10:1
>10:1
>10:1
>10:1

250 kHz Filter:

10.2375V
-10.24V

10.1863

-10.2912

____________
____________

10.2887

-10.1888

4.4E-5
2.3E-5

>10:1
>10:1

10 MHz Filter:

10.2375V
-10.24V

10.1863

-10.2912

____________
____________

10.2887

-10.1888

4.4E-5
2.3E-5

>10:1
>10:1

Test 2-3. DC Offset Test (Values in Vdc)

13 dB attenuation:

9.755V
4.0V
-4.0V
-9.755V

9.6355
3.9380

-4.0620
-9.8746

____________
____________
____________
____________

9.8746
4.0620

-3.9380
-9.6355

4.7E-5
2.04E-5
2.04E-5
4.7E-5

>10:1
>10:1
>10:1
>10:1

28 dB attenuation:

2.0V
-2.0V

1.9756

-2.0244

____________
____________

2.0244

-1.9756

1.12E-5
1.12E-5

>10:1
>10:1

Test 2-4. AC Accuracy Test (Values in Vac)

No Filter:

7.239V (0 dB atten)
6.45V (1 dB atten)
5.75V (2 dB atten)
4.566V (4 dB atten)
2.881V (8 dB atten)
1.440V (14 dB atten)
0.229V (30 dB atten)

7.1561
6.3396
5.6516
4.4878
2.8317
1.4153
0.2251

____________
____________
____________
____________
____________
____________
____________

7.3228
6.5624
5.8502
4.6455
2.9312
1.4651
0.2330

2.46E-3
2.3E-3
2.16E-3
1.9E-3
1.58E-3
1.29E-3
1.46E-4

>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1

250 kHz Filter:

7.239V (0 dB atten)

7.1561

____________

7.3228

2.46E-3

>10:1

78 Verification Tests

Agilent E1445A Service Manual

Advertising
This manual is related to the following products: