Test 2-1: dc zeros – Agilent Technologies 75000 Series C User Manual

Page 35

Advertising
background image

Test 2-1: DC Zeros

Description

The purpose of this test is to verify that the AFG meets its specifications for
DCV accuracy for an output of zero volts. An arbitrary waveform
consisting of zeros is used. The amplitude is varied in order to test each
attenuator.

Equipment Setup

Connect equipment as shown in Figure 2-5

Set DMM to: DCV, 100 mV range

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Delete all sequences and segments from memory:

LIST:SSEQ:DEL:ALL

Delete all sequences

LIST:SEGM:DEL:ALL

Delete all segments

Figure 2-5. Equipment Setup for Test 2-1 thru Test 2-4

Agilent E1445A Service Manual

Verification Tests 33