Test 2-3: dc offset (cont’d) – Agilent Technologies 75000 Series C User Manual

Page 46

Advertising
background image

Test 2-3: DC Offset (cont’d)

Example Program (cont’d)

440 FOR I=1 TO 6
450 IF I<=4 THEN
460 Vout=2.2919
470 ELSE
480 Vout=.40756
490 END IF
500 !
510 IF Vout<>Vout_old THEN
520 !Set offset to zero before changing amplitude
530 OUTPUT @Afg;":VOLT:OFFS 0;";
540 OUTPUT @Afg;":VOLT "&VAL$(Vout)&";";
550 END IF
560 !
570 OUTPUT @Afg;":VOLT:OFFS "&VAL$(Offset(I))!Set offset
580 PRINT Vout,Offset(I)
590 !
600 DISP "Record DMM reading, then press ’Continue’"
610 PAUSE
620 DISP
630 Vout_old=Vout
640 NEXT I

!Next attenuation

650 !
660 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

670 END
680 !
690 SUB Def_seq_zeros
700 COM @Afg
710 OUTPUT @Afg;"LIST:SEGM:SEL ZEROS"

!Segment name

720 OUTPUT @Afg;"LIST:SEGM:DEF 8"

!Segment length

730 OUTPUT @Afg;"LIST:SEGM:VOLT 0,0,0,0,0,0,0,0"

!Voltage points

740 !
750 OUTPUT @Afg;"LIST:SSEQ:SEL DC_ZEROS"

!Sequence name

760 OUTPUT @Afg;"LIST:SSEQ:DEF 1"

!# of segments

770 OUTPUT @Afg;"LIST:SSEQ:SEQ ZEROS"

!Segment list

780 SUBEND

44 Verification Tests

Agilent E1445A Service Manual