Test 2-1: DC Zeros (cont’d)

Example Program

This program performs the DC Zeros test. An arbitrary waveform, consisting of zeros, is used with various
amplitudes to test a variety of attenuator and filter combinations.

10! RE-STORE "DC_ZEROS"
20 COM @Afg
30 DIM Attn(1:9),Vout(1:9)
40 !
50 !---------- Set up I/O path and reset AFG ----------
60 ASSIGN @Afg TO 70910
70 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

80 !
90 !---------- Initialize variables ----------
100 DATA 0,.99,1,2,4,8,13,14,30
110 READ Attn(*)

!Read in attenuations

120 !
130 DATA 10.2375,9.13469,9.12416,8.13192,6.45941,4.0756
140 DATA 2.29187,2.04263,0.32372
150 READ Vout(*)
160 !
170 !---------- Set up DMM ----------
180 PRINT "Set up DMM:"
190 PRINT
200 PRINT " Function -- DCV"
210 PRINT " Range -- 100 mV"
220 PRINT
230 PRINT "Connect DMM HI and LO to AFG Output."
240 DISP "Press ’Continue’ when ready"
250 PAUSE
260 CLEAR SCREEN
270 !
280 !---------- Set up AFG ----------
290 OUTPUT @Afg;"*RST"

!Reset AFG

300 OUTPUT @Afg;"LIST:SSEQ:DEL:ALL"

!Delete all sequences

310 OUTPUT @Afg;"LIST:SEGM:DEL:ALL"

!Delete all segments

320 WAIT .5
330 OUTPUT @Afg;"ROSC:SOUR CLK10;";

!10MHZ clock

340 OUTPUT @Afg;":VOLT MAX;";

!MAX output

350 OUTPUT @Afg;":OUTP:LOAD INF;";

!Infinite load

360 OUTPUT @Afg;":FUNC USER"

!User waveform

370 !
380 CALL Def_seq_zeros

!Define waveform

390 OUTPUT @Afg;"FUNC:USER DC_ZEROS"

!Select sequence

400 OUTPUT @Afg;"INIT:IMM"
410 !
420 !---------- Perform test ----------
430 PRINT "ATTEN","FILTER","AMPLITUDE"

(Continued on next page)