Test 2-2: dc accuracy – Agilent Technologies 75000 Series C User Manual

Page 40

Advertising
background image

Test 2-2: DC Accuracy

Description

The purpose of this test is to verify that the AFG meets its specifications for
DC accuracy.

Equipment Setup

Connect equipment as shown in Figure 2-5

Set DMM to DCV, autorange

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Set up the AFG to output a DC signal:

FUNC DC;

Select DC waveform

:OUTP:LOAD INF;

Infinite load

:VOLT MAX

Set amplitude

Perform steps 3 - 5 for each amplitude listed in Table 2-2:

3.

Set up the AFG output filter as specified in Table 2-2. Use the
appropriate command(s) below:

OUTP:FILT OFF

Disable filter

or

OUTP:FILT:FREQ 250KHZ

Select 250 kHz filter

OUTP:FILT ON

Enable filter

or

OUTP:FILT:FREQ 10MHZ

Select 10 MHz filter

OUTP:FILT ON

Enable filter

4.

Set the AFG output amplitude:

VOLT <amplitude>

Set amplitude

where <amplitude> is the value specified in Table 2-2.

5.

Trigger the DMM and record the reading.

38 Verification Tests

Agilent E1445A Service Manual