Test 2-1: dc zeros (cont’d) – Agilent Technologies 75000 Series C User Manual

Page 39

Advertising
background image

Test 2-1: DC Zeros (cont’d)

Example Program (cont’d)

440 PRINT
460 FOR Filter=0 TO 2
470 SELECT Filter
480 CASE 0

!No filter

490 OUTPUT @Afg;"OUTP:FILT OFF"
500 Filter$="NONE"
510 CASE 1

!250KHZ filter

520 OUTPUT @Afg;"OUTP:FILT:FREQ 250KHZ"
530 OUTPUT @Afg;"OUTP:FILT ON"
540 Filter$="250 kHz"
550 CASE 2

!10MHZ filter

560 OUTPUT @Afg;"OUTP:FILT:FREQ 10MHZ"
570 OUTPUT @Afg;"OUTP:FILT ON"
580 Filter$="10 MHz"
590 END SELECT
600 !
610 FOR I=1 TO 9

!Loop through atten’s

620 OUTPUT @Afg;":VOLT "&VAL$(Vout(I))

!Set AFG amplitude

630 PRINT Attn(I),Filter$,Vout(I)
640 !
650 DISP "Record DMM reading, then press ’Continue’"
660 PAUSE
670 DISP
680 NEXT I

!Next attenuation

690 PRINT
700 NEXT Filter

!Next filter

710 !
720 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

730 END
740 !
750 SUB Def_seq_zeros
760 COM @Afg
770 OUTPUT @Afg;"LIST:SEGM:SEL ZEROS"

!Segment name

780 OUTPUT @Afg;"LIST:SEGM:DEF 8"

!Segment length

790 OUTPUT @Afg;"LIST:SEGM:VOLT 0,0,0,0,0,0,0,0"

!Voltage points

800 !
810 OUTPUT @Afg;"LIST:SSEQ:SEL DC_ZEROS"

!Sequence name

820 OUTPUT @Afg;"LIST:SSEQ:DEF 1"

!# of segments

830 OUTPUT @Afg;"LIST:SSEQ:SEQ ZEROS"

!Segment list

840 SUBEND

Agilent E1445A Service Manual