Test 2-3: dc offset (cont’d) – Agilent Technologies 75000 Series C User Manual

Page 44

Advertising
background image

Test 2-3: DC Offset (cont’d)

Test Procedure
(cont’d)

Perform steps 5 - 7 for each offset listed in Table 2-3:

5.

If necessary, change the AFG output amplitude:

VOLT:OFFS 0;

Set offset to 0

:VOLT <amplitude>

Set amplitude

where <amplitude> is the value specified in Table 2-3.

6.

Set AFG offset voltage:

VOLT:OFFS <offset>

Set offset

where <offset> is the value specified in Table 2-3.

7.

Trigger the DMM and record the reading.

Table 2-3. DC Offset Test Points

Offset

(volts)

Amplitude

(volts)

Test Limits

(volts)

9.755
4.000

-4.000
-9.755

2.000

-2.000

2.29189
2.29189
2.29189
2.29189
0.40756
0.40756

9.755

±

0.1196

4.0

±

0.0620

-4.0

±

0.0620

-9.755

±

0.1196

2.0

±

0.0244

-2.0

±

0.0244

42 Verification Tests

Agilent E1445A Service Manual