Test 2-7: Frequency Accuracy

Description

The purpose of this test is to verify that the AFG meets its specifications for
frequency accuracy.

Equipment Setup

Connect equipment as shown in Figure 2-8

Set Counter to: Frequency, 50

input impedance

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

Perform steps 2 - 6 for each entry listed in Table 2-7:

2.

Abort the waveform if it has been previously initiated:

ABORT

Figure 2-8. Equipment Setup for Test 2-7

56 Verification Tests

Agilent E1445A Service Manual