Test 2-1: dc zeros (cont’d) – Agilent Technologies 75000 Series C User Manual

Page 37

Advertising
background image

Test 2-1: DC Zeros (cont’d)

Test Procedure
(cont’d)

Table 2-1. DC Zeros Test Points

Attenuation

(dB)

Amplitude

(volts)

Filter

Test Limits

(volts)

0

.99

1
2
4
8

13
14
30

10.23750

9.13469
9.12416
8.13192
6.45941
4.07560
2.29187
2.04263
0.32372

None
None
None
None
None
None
None
None
None

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0044

0

±

0.0044

0

.99

1
2
4
8

13
14
30

10.23750

9.13469
9.12416
8.13192
6.45941
4.07560
2.29187
2.04263
0.32372

250 kHz
250 kHz
250 kHz
250 kHz
250 kHz
250 kHz
250 kHz
250 kHz
250 kHz

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0044

0

±

0.0044

0

.99

1
2
4
8

13
14
30

10.23750

9.13469
9.12416
8.13192
6.45941
4.07560
2.29187
2.04263
0.32372

10 MHz
10 MHz
10 MHz
10 MHz
10 MHz
10 MHz
10 MHz
10 MHz
10 MHz

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0220

0

±

0.0044

0

±

0.0044