Skew DAC Adjustment Procedure (cont’d)

Adjustment Procedure (cont’d)

2.

Set up the AFG to output an 11 dBm, 4 MHz sinewave:

FUNC SIN;
:VOLT 11 DBM;
:FREQ 4E6
INIT:IMM

3.

Load an initial value of 128 into the delay DAC:

DIAG:POKE #HE0000B,8,2
DIAG:POKE #HE0000D,8,128
DIAG:POKE #HE0000B,8,7
DIAG:POKE #HE0000D,8,8

4.

With the Spectrum Analyzer, locate and center the second harmonic.
Then, reduce the frequency span to 2 kHz.

5.

Find the delay DAC setting that minimizes the amplitude of the
second harmonic (see the example program).

6.

Disable calibration security on the AFG:

CAL:SEC:STAT OFF, <security code>

Cal security off

where <code> is the AFG’s security code (factory-set to "E1445A").

7.

Transfer the calibration constant (DAC setting) to the AFG in
arbitrary block data format:

CAL:DATA:SKEW <data >

Transfer cal constant

NOTE

See the example program to see how step 7 is performed in BASIC.

8.

Enable calibration security on the AFG:

CAL:SEC:STAT ON

Cal security on

110 Adjustments

Agilent E1445A Service Manual