Test 2-10: Spurious/Non-Harmonic Distortion (cont’d)

Example Program

This program performs the Spurious/Non-harmonic Test.

10 ! RE-STORE "NON_HARM"
20 DIM Start_freq(1:9),Stop_freq(1:9),Max_ampl(1:9)
30 !
40 !---------- Set up I/O path and reset AFG ----------
50 ASSIGN @Afg TO 70910

!AFG I/O path

60 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

70 !
80 !---------- Initialize variables ----------
90 Freq_out=1.0E+7

!Freq = 10 MHz

100 Dbm_out$="-5DBM"

!Amplitude = -5dBm

110 !
120 DATA 100E3,10.5E6,21E6,31E6,41E6,51E6,75E6,100E6,125E6
130 READ Start_freq(*)

!Read start freqs

140 !
150 DATA 9.5E6,19E6,29E6,39E6,49E6,75E6,100E6,125E6,150E6
160 READ Stop_freq(*)

!Read stop freqs

170 !
180 !---------- Set up Spec Analyzer ----------
190 CLEAR SCREEN
200 PRINT "Set up Spectrum Analyzer:"
210 PRINT
220 PRINT " Ref Level = -5dBm"
230 PRINT " Resolution BW = 3 kHz"
240 PRINT " Video BW = 3 kHz"
250 PRINT
260 PRINT "Connect Spectrum Analyzer to AFG Output."
270 DISP "Press ’Continue’ when ready"
280 PAUSE
290 CLEAR SCREEN
300 !
310 !---------- Set up AFG ----------
320 OUTPUT @Afg;"*RST"

!Reset AFG

330 WAIT 1
340 OUTPUT @Afg;"FREQ "&VAL$(Freq_out)&";";

!Set frequency

350 OUTPUT @Afg;":VOLT "&Dbm_out$

!Set amplitude

360 OUTPUT @Afg;"OUTP:FILT:FREQ 10MHZ"

!Enable 10MHz filter

370 OUTPUT @Afg;"OUTP:FILT ON"
380 OUTPUT @Afg;"INIT:IMM"

!Initiate

390 !

(Continued on next page)

Agilent E1445A Service Manual

Verification Tests 71