Functional Verification: Self-Test

Description

The AFG self-test performs the following internal checks:

internal interrupt lines

waveform select RAM

segment sequence RAM

waveform segment RAM

DDS/NCO operation

sine wave generation

arbitrary waveform generation

marker generation

waveform cycle and arm counters

sweep timer

frequency-shift keying

stop trigger

DC analog parameters (amplitude, offset, attenuators, filters,
calibration DACs)

Test Procedure

1.

Remove any connections to the AFG front panel.

2.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

3.

Execute the AFG self-test:

*TST?

Self-test command

4.

Read the result. A "0" indicates that the test passed. A "1" indicates a
failure. Read the error queue using the SYST:ERR? command until
the error message is "No error".

Agilent E1445A Service Manual

Verification Tests 19