Functional Verification: Ref In/Marker Out Test

Description

The purpose of this test is to check the Ref/Sample In and Marker Out ports.
An external reference is connected to the Ref/Sample In port and sent to the
Marker Out port.

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Set up equipment as shown in Figure 2-1:

3.

Set up the AFG to output the external reference to the "Marker Out"
port:

ROSC:SOUR EXT

External ref oscillator

MARK:FEED "ROSC"

Marker source is ROSC

INIT:IMM

Initiate

4.

Verify that the scope shows a 10 MHz squarewave.

Figure 2-1. Ref/Sample In Test Setup

20 Verification Tests

Agilent E1445A Service Manual

Advertising
This manual is related to the following products: