Test 2-4: AC Accuracy (cont’d)

Example Program (cont’d)

420 FOR I=1 TO 9
430 SELECT Filter(I)
440 CASE 0
450 OUTPUT @Afg;"OUTP:FILT OFF"

!No filter

460 Filter$="NONE"
470 CASE 1
480 OUTPUT @Afg;"OUTP:FILT:FREQ 250KHZ"

!250 kHz filter

490 OUTPUT @Afg;"OUTP:FILT ON"
500 Filter$="250 kHz"
510 CASE 2
520 OUTPUT @Afg;"OUTP:FILT:FREQ 10MHZ"

!10 MHz filter

530 OUTPUT @Afg;"OUTP:FILT ON"
540 Filter$="10 MHz"
550 END SELECT
560 OUTPUT @Afg;":VOLT "&VAL$(Vout(I))&"VRMS"

!Set amplitude

570 PRINT Filter$,Vout(I)
580 WAIT .5
590 !
600 DISP "Record DMM reading, then press ’Continue’"
610 PAUSE
620 DISP
630 NEXT I
640 !
650 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

660 END

48 Verification Tests

Agilent E1445A Service Manual