Test 2-2: dc accuracy (cont’d) – Agilent Technologies 75000 Series C User Manual

Page 41

Advertising
background image

Test 2-2: DC Accuracy (cont’d)

Test Procedure
(cont’d)

Table 2-2. DC Accuracy Test Points

Amplitude

(volts)

Filter

Test Limits

(volts)

10.2375

5.0
0.0

-5.0

-10.24

10.2375

-10.24

10.2375

-10.24

None
None
None
None
None

250 kHz
250 kHz

10 MHz
10 MHz

10.2375

±

0.0512

5.0

±

0.0355

0.0

±

0.0205

-5.0

±

0.0355

-10.24

±

0.0512

10.2375

±

0.0512

-10.24

±

0.0512

10.2375

±

0.0512

10.24

±

0.0512

Example Program

This program performs the DC Accuracy test.

10! RE-STORE "DC_LEVELS"
20 DIM Vout(1:9),Filter(1:9)
30 !
40 !---------- Set up I/O path and reset AFG -----------
50 ASSIGN @Afg TO 70910
60 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

70 !
80 !---------- Initialize variables ----------
90 DATA 10.2375,5.0,0,-5.0,-10.24,10.2375,-10.24,10.2375,-10.24
100 READ Vout(*)
110 !
120 DATA 0,0,0,0,0,1,1,2,2
130 READ Filter(*)
140 !
150 !---------- Set up DMM ----------
160 CLEAR SCREEN
170 PRINT "Set up DMM:"
180 PRINT
190 PRINT " Function -- DCV"
200 PRINT " Range -- AUTO"
210 PRINT
220 PRINT "Connect DMM HI and LO to AFG Output."
230 DISP "Press ’Continue’ when ready"
240 PAUSE
250 CLEAR SCREEN

Agilent E1445A Service Manual

Verification Tests 39

Advertising
This manual is related to the following products: