Test 2-7: Frequency Accuracy (cont’d)

Test Procedure
(cont’d)

Table 2-7. Frequency Accuracy Test Points

Ref Oscillator

Source

Marker

Source

Squarewave

Frequency (Hz)

Test Limits

(Hz)*

INT1
INT2
INT2
INT2
INT2

"ROSC"
"ROSC"
"TRIG"
"TRIG"
"TRIG"

----------
----------
5.0 E6
3.333 E3
76.294

42.94967 E6

±

0.005%

40 E6

±

0.005%

20 E6

±

0.005%

13.3333 E6

±

0.005%

305.176

±

0.005%

*Add aging rate of

±20 ppm/year

Example Program

This program performs the Frequency Accuracy Test.

10! RE-STORE "OSC_FREQ"
20 DIM Freq(1:5)
30 !
40 !---------- Set up I/O path and reset AFG ----------
50 ASSIGN @Afg TO 70910
60 OUTPUT @Afg;"*RST;*CLS"

!Reset AFG

70 !
80 !---------- Initialize variables ----------
90 DATA 42.94967E6,40E6,20E6,13.3333E6,305.176
100 READ Freq(*)
110 !
120 !---------- Set up Counter ----------
130 CLEAR SCREEN
140 PRINT "Set up Counter:"
150 PRINT
160 PRINT " Function -- Frequency"
170 PRINT " Input Impedance -- 50 ohms"
180 PRINT
190 PRINT "Connect the Counter to ’Marker Out’ on the E1445A."
200 PRINT
210 DISP "Press ’Continue’"
220 PAUSE
230 CLEAR SCREEN
240 !

(Continued on next page)

58 Verification Tests

Agilent E1445A Service Manual