DC Adjustment Procedure (cont’d)

Test Procedure
(cont’d)

5.

Trigger the DMM and note the reading.

6.

Send the reading to the AFG:

CAL:DC:POINT? <reading>

where <reading> is the DMM reading from step 5. The AFG will
return, in order, the number of the current cal point and an error code.
Any non-zero error code indicates a failure.

Table 3-1. DMM Range Setting for Cal Points 41 and 43

Absolute Value

of DC Offset

DMM Range

Less than 10 mV

100

Between 10 mV and 100 mV

1000

Greater than 100 mV

Offset too high - abort cal

Table 3-2. DC Calibration Points

Cal

Point

DMM Settings

(changes only)

1
29
31
33
41
43

DCV, 10 V range, 10 NPLC
1 V range, 20 NPLC
10 V*
100 mV range
4-wire ohms, offset comp on*
4-wire ohms, offset comp on*

*See Step 4 of the "DC Adjustment Procedure" for
special instructions.

Agilent E1445A Service Manual

Adjustments 89