Functional Verification: Start Arm In Test (cont’d)

Test Procedure
(cont’d)

4.

Set up the AFG to output a 1 MHz sinewave, with an external Start
Arm source:

FREQ 1E6;

Set freq to 1 MHz

:VOLT 4VPP

Set AFG amplitude

ARM:LAY2:SOUR EXT

External Start Arm
source

INIT:IMM

Initiate

5.

Verify that no signal appears on the scope. Send the following
command to the Command Module to provide a Start Arm signal to
the AFG:

OUTP:EXT:LEV 0

6.

Verify that a 1 MHz sinewave appears on the scope.

22 Verification Tests

Agilent E1445A Service Manual