Test 2-8: Duty Cycle

Description

The purpose of this test is to verify that the AFG meets its specifications for
square wave duty cycle. Duty cycle is determined by measuring positive
pulse width.

Equipment Setup

Connect equipment as shown in Figure 2-9

Set Counter to: Pulse Width, DC coupling, 50

input impedance

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Set the AFG to output a square wave:

FUNC SQU;

Select squarewave

:VOLT MAX

Set to max amplitude

Perform steps 3 - 7 for each frequency listed in Table 2-8:

3.

Abort the waveform if it has been previously initiated:

ABORT

Figure 2-9. Equipment Setup for Test 2-8

60 Verification Tests

Agilent E1445A Service Manual