5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1028

■■■■■■■■

tiamo 2.3

5.6.4.11.3

MEAS TC Cond - Measuring parameters

Tab: Method

MEAS TC Cond Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement frequency

712
Selection

auto | 300 Hz | 2.4 kHz

Default value

auto

auto
The more suitable of the two frequencies is selected.

Stop time

Maximum period of time for measurement.

Range

0 ... 999999 s

Default value

3000 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

712
Range

0.4 ... 999999 s

Default value

2.0 s

856
Range

0.1 ... 999999 s

Default value

2.0 s

Sample solution

Selection of the sample solution from the table Sample solutions (TC
conductivity)
.

Selection Table entries
Default value first entry of table

Start temperature

Temperature at the start of the measurement.

Range

-20.0 ... 150.0 °C

Default value

20.0 °C