11 met ipol - fixed endpoint evaluation – Metrohm tiamo 2.3 Manual User Manual

Page 659

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3

■■■■■■■■

643

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.3.4.11

MET Ipol - Fixed endpoint evaluation

Dialog window: Method

MET Ipol Properties... Additional evalua-

tions

Fixed endpoint evaluation [New]/[Properties] Fixed endpoint evalua-

tion #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL