11 met u - fixed endpoint evaluation, 4 met ipol, 1 met ipol - overview – Metrohm tiamo 2.3 Manual User Manual

Page 633

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3

■■■■■■■■

617

5.6.3.3.3.11

MET U - Fixed endpoint evaluation

Dialog window: Method

MET U Properties... Additional evaluations Fixed

endpoint evaluation

[New]/[Properties] Fixed endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.3.4

MET Ipol

5.6.3.3.4.1

MET Ipol - Overview

Dialog window: Method

MET Ipol Properties... MET Ipol - 'Command name'

Command for Monotonic equivalence point titrations with voltamet-
ric measurement (selectable polarization current).

Instruments
This command can be executed with the following devices.

Titrando: 808, 809, 835, 836, 852, 857, 888, 904, 905, 906, 907

Titrino: 702, 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance: