3 live display - application note, 4 live display - messages, 5 live display - properties – Metrohm tiamo 2.3 Manual User Manual

Page 196: Live display - application note, Live display - messages, Live display - properties

Advertising
background image

3.8 Subwindow Live display

■■■■■■■■■■■■■■■■■■■■■■

180

■■■■■■■■

tiamo 2.3

Cancel

Cancel the active command of the track.

Properties - Live dis-
play

Set properties of the live display (see Chapter 3.8.5, page 180).

3.8.3

Live display - Application note

Subwindow: Workplace

Live display

On the tab Application note, the application note defined in the START
command of the loaded method is displayed. This tab is opened by
default in the subwindow Live display when loading the method.

3.8.4

Live display - Messages

Subwindow: Workplace

Live display

On the tab Messages there is a scrollable message field which contains
all messages concerning events occurring during the run of the current
determination. Entries consist of messages that generate a database entry
in the determination but which are not significant enough to result in the
run being interrupted pending confirmation by the user.

Each message consists of Date, Time and Message text.

The message window is deleted when a run is started with [START]. In
this way all messages of the last single determination or determination
series can be viewed.

The label of the tab changes to red as soon as a new message has been
entered. The label of the Messages tab changes to black again after
having shown the messages and the tab is changed anew.

3.8.5

Live display - Properties

Subwindow: Workplace

Live display

With the menu item View, Properties, Properties Live display # or
the context-sensitive menu item Properties Live display # in the sub-
window itself the dialog window Properties Live display # is opened.

Command type

Selection of the command type for which the properties in the live display
are to be defined. The curve properties defined for each command type
are saved per live display window and per client.

Selection

DET | MET | SET | MEAS | MEAS T/Flow | MEAS
TC Cond
| MEAS Opt | MEAS TMF | KFT | KFC /
BRC
| STAT | DOS | TET

Default value

DET