5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

840

■■■■■■■■

tiamo 2.3

Selection

Measured value | Time | Water

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Water
Range

0.0 ... 999999.9 µg

5.6.3.7

BRC

5.6.3.7.1

BRC - Overview

Dialog window: Method

BRC Properties... BRC - 'Command name'

Command for coulometric titration for determining the bromine con-
sumption. Out of the amount of bromine consumed the bromine index
(amount of bromine consumed (in mg) per 100 g sample) and the bro-
mine number
(amount of bromine (in g) per 100 g sample) can be calcu-
lated.

Instruments
This command can be executed with the following devices.

Titrando: 851, 852

Appearance
The command has the following appearance:

without conditioning

with conditioning

Parameters
The parameters for the command BRC are configured in the following 8
tabs:

General/Hardware
Parameters for devices, indicator and generator electrode and stirrers.

Start conditions
Entry of a waiting period before the start of the titration.