5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1170

■■■■■■■■

tiamo 2.3

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Range

0.1 ... 999999 s

Default value

5.0 s

Titrino
Range

1 ... 999999 s

Default value

5 s

5.6.6.3.4.4

DOS pH - Stop conditions

Tab: Method

DOS pH Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the dosing are defined on this tab. If more
than one stop condition is in effect at the same time, then the criterion
which is fulfilled first will stop the dosing.

Stop volume

Stops when the given volume has been added after the start of the dos-
ing. The stop volume should be adapted to suit the sample weight or the
titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100.000 mL

Selection

off

off
Means that no stop will take place.

Titrino
Range

0 ... 9999.99 mL

Default value

100.00 mL