5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1168

■■■■■■■■

tiamo 2.3

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 855
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos and 855.

Switch on/off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command. This parameter is displayed only for Titrinos.

5.6.6.3.4.3

DOS pH - Dosing parameters

Tab: Method

DOS pH Properties... Dosing parameters

Command name

Name of the command.

Input

25 characters

The volume to be added, the dosing rate and the dosing time are defined
on this tab.

Dosing criterion

Selects the two parameters that can be preset for dosing. The third para-
meter is calculated automatically from the formula Volume = dosing
time * dosing rate
and is not shown (see Chapter 5.6.6.3.2, page
1157)
.

Selection

Volume/dosing rate | Volume/dosing time |
Dosing rate/dosing time

Default value

Volume/dosing rate

Volume

Fixed volume that is to be dosed. This parameter is only displayed for
Dosing criterion = Volume/Dosing rate or Volume/Dosing time.