5 Method

tiamo 2.3

■■■■■■■■

853

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the
subwindow Live display of the program part Workplace, in addition to
the drift and the time.

5.6.3.7.8

BRC - Additional evaluations

Tab: Method

BRC Properties... Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional method for evaluation of the titration curves can
be activated and defined on this tab:

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fixed measured quantity (meas-
ured value
, time or bromine) for the fixed endpoint from the measur-
ing point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.7.10, page 855).

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.7.10, page 855).

[Delete]

Delete the selected line.