6.3 Configuration data

■■■■■■■■■■■■■■■■■■■■■■

1328

■■■■■■■■

tiamo 2.3

6.3.3.2.2

Editing templates for input lines

Dialog window: Configuration

Tools Templates Input lines Templates for

input lines

[New] / [Edit] New template / Edit template

With [New] or [Edit] the dialog window New template or Edit tem-
plate
opens for entering a new template or for editing an existing tem-
plate.

Signal name

Name of the pattern for the input signal.

Input

25 characters

Input signal

Entry of the bit pattern for the input signal with exactly 8 characters. It is
possible to enter the characters

0 = Line inactive,

1 = Line active and

* = Any line status.

Selection

Bit pattern with 8 characters (0, 1, *) | ********

Default value

********

The input lines and bits are numbered from right to left:

Input 7 6 5 4 3 2 1 0

Bit

7 6 5 4 3 2 1 0

Example: *******1 expects an active input line 0 (1 = set). For example,
this line is set by a Titrino after a titration has been completed and the
Titrino can accept a start signal again.

Note

Input lines that are of no interest or for which no defined condition can
be predicted should also be masked with an asterisk *.

6.3.3.3

Templates - Output lines

6.3.3.3.1

Managing templates for output lines

Dialog window: Configuration

Tools Templates Output lines... Templates

for output lines

The client-specific bit pattern for setting remote output signals can be
defined in the dialog window Templates for output lines; these can be
selected with the command CTRL and in Manual Control. The table with
the defined templates cannot be edited, although it can be sorted accord-