6.7 Subwindow Sensors

■■■■■■■■■■■■■■■■■■■■■■

1406

■■■■■■■■

tiamo 2.3

6.7.3.7

Initial sensor data

Dialog window: Configuration

Sensors Edit Properties Calibration

data

Initial data sensor

The initial calibration data for the intelligent sensor is displayed in the dia-
log window Initial sensor data.

Slope

Display of the slope in % for the first calibration.

U

as

Display of the asymmetry potential U

as

in mV during the first calibration.

Calibration temperature

Display of the temperature in °C during the first calibration.

6.7.3.8

Sensor history - Limits

Dialog window: Configuration

Sensors Edit Properties History [Limit val-

ues]

Limits Sensor

In the dialog window Limits for sensor you can define warning and
intervention limits for the slope and pH(0) or E(0). If you have defined lim-
its then these will be shown in the graph in orange for warning limits and
red for intervention limits. However, whether these limits are observed is
not monitored, i.e. exceeding these limits does not trigger any action. For
monitoring during calibration, the limits set on the Limits tab apply.

Warning limits for slope (for pH and ISE electrodes only)
on | off (Default value: off)

If this option is activated, then the values for the slope will be shown in
orange if the limits are exceeded.

Lower limit

Lower warning limit for the slope.

for pH electrode
Range

-2.147E+9 ... 2.147E+9 %

Default value

97.5 %

for ISE (+1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

57.2 mV

for ISE (-1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-61.2 mV