6.3 Configuration data

■■■■■■■■■■■■■■■■■■■■■■

1330

■■■■■■■■

tiamo 2.3

Example: ************1* sets the output line 1 to active (= set), which
would result in a Stop command for a connected Titrino, for example.
************0* sets the line to inactive.

Note

We recommend masking the irrelevant output lines with an asterisk *
so as not to modify these line conditions.

Pulse length

Duration of the transmitted pulse.

Range

100 ... 1000 ms

Default value

200 ms

Examples for 765 Dosimat

Parameters

Binary pattern

Function

Start Dos1

*******p******

Starts Dosimat 1 (Titrino via "acti-
vate")

Start Dos2

*****p********

Starts Dosimat 2 (Titrino via "acti-
vate")

Start Dos*

*****p*p******

Starts Dosimat 1 and 2 (Titrino via
"activate")

6.3.3.4

Templates - Conductivity standards

6.3.3.4.1

Manage conductivity standards

Dialog window: Configuration

Tools Templates Conductivity stand-

ards

Conductivity standards

Conductivity standards can be defined in the dialog window, which
can be selected with the command CAL Cond. The table cannot be edi-
ted.

[New]

Creation of a new conductivity standard. The window Edit conductivity
standard
opens, in which a new conductivity standard can be defined
(see Chapter 6.3.3.4.2, page 1331).

[Properties]

Edits the selected conductivity standard. The window Edit conductivity
standard
opens, in which the conductivity standard can be edited (see
Chapter 6.3.3.4.2, page 1331)
.