5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

728

■■■■■■■■

tiamo 2.3

Titrando, 719, 720, 736, 751, 758, 785, 794, 798, 799
Range

0 ... 999999 s

Temperature

Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware
tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously. This value is used for temperature
correction in pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Range

0.1 ... 999999 s

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s

Default value

2.00 s

5.6.3.4.4.6

SET U - Stop conditions

Tab: Method

SET U Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.