6 sample changer - heater/gas, Heater/gas – Metrohm tiamo 2.3 Manual User Manual

Page 1613

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Manual control

tiamo 2.3

■■■■■■■■

1597

Operation

Selection whether the pump or valve should be switched on/off manually
or whether they should be switched on for a defined duration and then
automatically be switched off.

Selection

On/Off | Duration

Default value

On/Off

Duration

Entry of the duration during which the pump/valve should remain
switched on. This field only appears for Operation = Duration.

Range

1 ... 999999 s

Default value

60 s

[Start]

Start pump 2. In the status display the running time since start is shown.

[Stop]

Stop pump 2.

8.4.4.6

Sample changer - Heater/Gas

Program part: Manual control

Here the heater and gas flow on the 774 and 874 Oven Sample Processor
can be switched on and off.

Heater
Heat the oven up to the desired temperature during the defined heating
up time.

Temperature

Temperature at which the oven has to be heated. With Init it is heated up
to the initial temperature set on the instrument.

Range

50 ... 250 °C

Default value

50 °C

Selection

Init

Heating time

Duration of the heating up period until reaching the desired temperature.

Range

1 ... 999 min

Selection

off

Default value

off

[Start]

Start the heater. In the status display the current temperature is shown.