6 properties - electrode test – Metrohm tiamo 2.3 Manual User Manual

Page 1420

Advertising
background image

6.7 Subwindow Sensors

■■■■■■■■■■■■■■■■■■■■■■

1404

■■■■■■■■

tiamo 2.3

[Limits]

Open the dialog window Limits for sensor, in which the warning and
intervention limits for the calibration data can be defined. These limits only
apply to the graph, no monitoring will take place during calibration. For
monitoring during calibration, the limits set on the Limits tab apply.

[Delete history]

Delete the sensor history.

6.7.3.6

Properties - Electrode test

Dialog window: Configuration

Sensors Edit Properties Electrode test

Date ELT

Date of the last electrode test. The date can be selected by pressing

in

the dialog window Select date. This date is used for the electrode test
monitoring (Date ELT + Interval ELT = Next ELT).

Result

Indicates the quality of the tested electrode.

Selection

Excellent | Good | Usable | Not tested

Default value

Not tested

Monitoring the ELT validity
on | off (Default value: off)

If this check box is activated, then the time interval for the electrode test is
monitored.

Can be edited only if a date has been entered in the field Date ELT.

Interval ELT

Time interval to next electrode test. If a value is entered here then the
date in the field Next ELT will be adjusted automatically.

Can be edited only for Monitor ELT validity = on, otherwise inactive.

Range

1 ... 999 Days

Default value

999 Days

Next ELT

Date on which the next electrode test is to be carried out. The date can be

selected by pressing

in the dialog window Select date. Once the date

has been entered, the ELT field will be adjusted automatically.

Can be edited only for Monitor ELT validity = on, otherwise inactive.

Default value

   Date ELT+ 999 days