6 Configuration

tiamo 2.3

■■■■■■■■

1375

Upper limit

Upper intervention limit.

Input

10 digits

Default value

1.0500

6.6.3.6

Properties - Exchange unit

Tab: Configuration

Titrants/solutions Properties Exchange unit

Hardware

Name

Freely definable designation for the exchange unit.

Input

24 characters

Default value

'empty'

Type

Display of the type of exchange unit:

Selection

EU | IEU

EU
Non-intelligent exchange unit without data chip.
IEU
Intelligent 806 Exchange Unit with data chip.

Order number

Order number of exchange unit. With intelligent exchange units the order
number is read off automatically and cannot be edited.

Input

24 characters

Default value

'empty'

Serial number

Serial number of exchange unit. With intelligent exchange units the serial
number is read off automatically and cannot be edited.

Input

10 characters

Default value

'empty'

Cylinder volume

Cylinder volume of exchange unit. With intelligent exchange units the cyl-
inder volume is read off automatically and cannot be edited. If in the
method you have selected a solution in a titration or dosing command
then the cylinder volume will be checked in the run.

Selection

1 mL | 5 mL | 10 mL | 20 mL | 50 mL

Default value

20 mL